EDPN IBAD-2612-ED Intelligent Battery Charge & Discharge Unit.
The IBAD-2612-ED Intelligent Battery Activator & Discharger has 3 independent test methods: battery discharge, battery charge and battery cycle.
Focusing on lagging single cells, it can run a discharge test to estimate the real capacity, charge the battery intelligently, and run multiple discharge/charge cycles to activate a lagging battery. Test data can be uploaded to a PC, where the analysis software helps identify the quality problem of the cell.
FEATURES
Features & benefits.
Features
1.2 V / 2 V / 6 V / 12 V charge & discharge: 1.2 V — 60 A discharge / 60 A charge; 2 V — 100 A / 100 A; 6 V — 30 A / 30 A; 12 V — 30 A / 30 A.
Multiple adjustable stop points to avoid excessive discharge and charge.
7 inch touch LCD displays all real-time data and graphs with direct click operation.
Non-volatile internal memory saves several independent discharge & charge records without a laptop; review, delete or download to PC via USB.
PC analysis software: battery voltage curve, current curve, capacity histogram and data form; automatic EXCEL data report.
Protection
Input over-voltage — LCD prompt + sound warning
Reverse polarity — LCD prompt + sound warning
Overload — LCD prompt + sound warning
Overheat — LCD prompt + sound warning
SPECIFICATIONS
Technical data.
EDPN product data. Confirm the current datasheet and the project specification before selection.
Discharge & charge current ranges
0.9 – 1.8 Vdc
Discharge 0 – 60 A / Charge 0 – 60 A
1.7 – 2.4 Vdc
Discharge 0 – 100 A / Charge 0 – 100 A
5.1 – 15 Vdc
Discharge 0 – 30 A / Charge 0 – 30 A
Current measurement
Current range
0 – 100 A
Accuracy & resolution
≤ ±1 %, 0.1 A
Voltage measurement
Voltage range
1.7 – 2.4 V
Accuracy & resolution
≤ ±1 %, 0.01 V
Voltage range
5.1 – 15 V
Accuracy & resolution
≤ ±0.25 %, 0.1 V
Power supply
Power supply voltage
220 VAC (-20 % to +30 %)
Frequency
50 / 60 Hz
Power consumption
350 W (max)
Communication & storage
Communication port
USB port
Internal memory
4 GB Flash
CE certificate
LVD standards
EN 61010-1:2010, EN 61010-2-030
EMC standards
EN 61326-1:2013, EN 61000-3-2:2014, EN 61000-3-3:2013
Environment
Operating temperature
0 to 45 °C
Storage temperature
-10 to 60 °C
Humidity
5 % – 95 % RH
Altitude
below 2000 m
Working noise
< 60 dB
Dimensions & weight
Main tester
348 x 222 x 256 mm
Carrying case
460 x 300 x 500 mm
Weight
10 kg (main tester) / 20 kg (with accessories and case)
DOWNLOADS
Datasheets & documents.
EDPN datasheet
The datasheet for this EDPN product is available on request.